Select another model: |
|
GTEM 1500 – Main Technical Specifications / Max EUT Size | |
Frequency range | 0,01 MHz – 20GHz* |
Septum Height [mm]: | 1500 |
Max EUT Size (LxWxH)cm | 110x110x100 cm |
Defined test Vol. within +6dB | 50x50x50 cm |
Typical VSWR | 1:1,2 |
Typical VSWR at the critical frequency | <1:1,6 |
Max Input power, W continuous/*pulsed | 700/*1,5Kw |
Input connector | N or 7/16″ |
Nominal impedance | 50 |
Window in the door | Diam.200 mm. |
Empty Technical panel | N.2 Diam.200 mm. |
Mechanical Dimensions | |
Outer LxWxH [cm]: | 710x358x255 |
Door WxH [cm]: | 80×120 *usefull passage. |
Weight Kg. Approx. | 1300 |
Wheeled undercarriage (Cost Option) | 25cm high |
Electrical Specifications | |
Mains connectors | Fix/CEE |
Main Switch | 16A magneto-thermal |
Input Socket plug | 16Aac (Single phase + ground) |
Output Socket EUT tape | 16Aac (Single phase + ground) |
Additional EUT sockets | Optional |
Ground connection | M6 bolt |
AC filter wires (Single PH+Ground) | 16Amp. 2 wires |
Channel for fiber optic leads | 3 couples |
RF feed-thru connectors | n.1 N, N.2 SMA double female |
Indoor LED illumination | n.2x20W |
Electrical Equipment / Options | |
AC filter 30A/2 wire (2PH+Ground) | |
AC filter 16A/4 wires (3PH+N+Ground) | |
AC/DC feed-thru filter 300A 450V | |
Electrical safety interlock | |
Multimedia filters: RJ45, 2xUSB, RS232 | |
25-poles signal filter (DB25) | |
Channel for fiber optic leads (6 couples) | |
Additional RF feed-thru N-type connector | |
Additional RF feed-thru SMA type connector | |
RJ11 (RJ9) feed-thru connector | |
Technical panel pre-drilled for options | |
Video camera system | |
Mechanical Equipment / Options | |
Gas / Water feedthrough plates | |
Honeycomb panel | |
Larger Door: 140x140cm. | |
Empty Technical panel
Secondary door:40x40cm or 60x60cm Manipulator |
Select another model: